LINQ II™ insertable cardiac monitor (ICM) System

MRI Conditions
for Use

Conditionally safe MRI access SureScan™ technology.
LINQ II ICM is MR Conditional at ≤ to 3.0 T with no post-insertion waiting required.

Healthcare professional smiling down at patient who is entering an MRI scanner surrounded by abstract blue dots and particles

A patient with a LINQ II device can be safely scanned in an MR system that meets the following conditions with no post-insertion waiting required. Failure to follow these conditions for use may result in a hazard to the patient during an MRI scan:

  • Open or closed clinical MRI systems with a static magnetic field of ≤ to 3.0 T must be used.
  • Hydrogen proton MRI equipment must be used.
  • Maximum spatial gradient of the static magnetic field specification must be ≤ 25 T/m (2500 gauss/cm).
  • Whole body gradient systems with gradient slew rate specification must be ≤ 200 T/m/s per axis.
  • The Whole Body Specific Absorption Rate (WB-SAR) as reported by the MRI equipment must be ≤ 4.0 W/kg; the head SAR as reported by the MRI equipment must be ≤ 3.2 W/kg.
  • Do not use local transmit coils on the chest, trunk, or shoulder region.
  • There are no restrictions on the placement of receive-only coils, and there are no restrictions on the use of local transmit or receive coils for imaging of the head or extremities.

For more information or to check MRI compatibility, visit the website below.