Features

Flexible anterior/posterior support

The unique composite design of the CG Future™ ring and band considers the full cardiac cycle.1 The ring or band provides enough stiffness to provide posterior remodeling while preserving dynamic annular motion.2–3

Long-term results of mitral valve repair with the CG Future™ band

“In the current era of mitral valve repair the gold standard should be both durability, in terms of freedom from reoperation and recurrent mitral insufficiency, and a physiologically normal valve without functional mitral stenosis (FMS) or systolic anterior motion (SAM). The design characteristics of a semirigid posterior annuloplasty band can increase the likelihood of achieving this goal.”3



Eight-year cumulative freedom
from adverse events with
the CG Future™ band and
competitive complete rings.3,4

retrospective analysis
of 1,612 patients with
degenerative mitral
regurgitation

View this chart to see long-term data for the CG Future™ annuloplasty ring and band.

† Non-randomized, non-matched comparison with statistically different baseline and procedural characteristics.

Key risks of mitral annuloplasty include recurrent regurgitation, stenosis, and ring/band dehiscence.

Designed for implantability

The CG Future™ ring and band can be used for small incisions where visibility and access to the anterior annulus may be limited.

The CG Future™ annuloplasty ring is used for mitral valve repair, designed to predictably remodel the annulus to maintain apposition of the anterior and posterior leaflets.

Low-profile design4

The two eyelets on the CG Future™ sizer align with the trigones for anchoring.

Two eyelets align with trigones for anchoring

Ordering information

CG Future™ annuloplasty ring, Model 638R

Item number Ring size (mm) A (mm) B (mm) C (mm) D (mm) Orifice area (mm2)
638RL24 24 15.27 16.53 24.0 28.9 269.3
638RL26 26 16.38 17.64 26.0 30.9 314.9
638RL28 28 17.48 18.73 28.0 32.9 364.3
638RL30 30 18.59 19.85 30.0 34.9 437.4
638RL32 32 19.71 20.97 32.0 36.9 474.7
638RL34 34 20.83 22.09 34.0 38.9 536.0
638RL36 36 21.92 23.18 36.0  40.9 600.9
638RL38 38 23.04 24.29 38.0 42.9 669.4

CG Future™ annuloplasty band, Model 638B

Item number Ring size (mm) A (mm) B (mm) C (mm) D (mm) Orifice area (mm2)
638BL26 26 16.38 17.64 26.0 30.9 332.0
638BL28 28 17.48 18.73 28.0 32.9 382.8
638BL30 30 18.59 19.85 30.0 34.9 437.4
638BL32 32 19.71 20.97 32.0 36.9 496.3
638BL34 34 20.83 22.09 34.0 38.9 559.2
638BL36 36 21.92 23.18 36.0 40.9 625.1
638BL38 38 23.04 24.29 38.0 42.9 695.7

CG Future™ accessories

Item number Accessories
7638S  CG Future™ sizer set (8 sizers) 
7615  Annuloplasty handle (216 mm length) 
7615XL  Annuloplasty handle (373 mm length) 
T7630 CG Future™ accessory tray 

Related links

TM* Third-party brands are trademarks of their respective owners. All other brands are trademarks of a Medtronic company.

† Non-randomized, non-matched comparison with statistically different baseline and procedural characteristics.

  1. Redmond J, Christiansen D, Bergin C, et al. In-vivo motion of mitral valve annuloplasty devices. J Heart Valve Dis. 2008;17(1):110–118.
  2. Sharony R, Saunders P, Nayar A, et al. Semirigid partial annuloplasty band allows dynamic mitral annular motion and minimizes valvular gradients: an echocardiographic study. Ann Thorac Surg. 2004;77:518–22.
  3. James L, Grossi EA, Loulmet DF, Galloway AC. Semirigid posterior annuloplasty band: Reshaping the mitral orifice while preserving its physiology. JTCVS Tech. 2021:37–42.
  4. Yaffee D, Loulmet D, Zias E, et al. Long-term results of mitral valve repair with semi-rigid posterior band annuloplasty. J Heart Valve Dis. 2014;23(1):66–71.