Portable patient monitoring

Nellcor™ portable SpO₂ patient monitor

<p>The Nellcor™ portable SpO₂ patient monitor is a convenient, handheld pulse oximeter ideal for spot-checks or continuous monitoring.</p>

Features

  • Small, lightweight, ergonomic handheld monitor for continuous and spot-check pulse oximetry monitoring
  • Supports a robust data set, including real-time SpO2 and pulse rate values, SatSeconds alarm management, pleth waveform, blip bar, and tabular trend data
  • Supports sleep study mode, which enables the clinician to dim the LCD display and silence alarms to prevent disruption of the patients’ sleep
  • Compatible with the complete line of Nellcor™ single-patient use, reusable and specialty sensors with OxiMax™ technology
  • 80 hours of data storage capability
  • Supports data export to an external personal computer for data analysis and printing functions
  • Supports standard and home-care modes for use in hospital, hospital-type facilities, transport, mobile environments, and home-care environments

Note: Transport and mobile environments are defined as intra hospital transport or transport between facilities or between a facility and a home environment

Clinical applications

Providing accurate, reliable SpO2 and pulse rate readings, this portable, easy-to-use device is an ideal solution for these critical screenings.

Specifications

Enclosure

Weight 274 g (0.604 lbs), including four batteries
Dimensions 70 mm W × 156 mm H × 32 mm D
(2.76 in W × 6.14 in H × 1.26 in D)
Display
Screen size 88.9 mm (3.5 in), measured diagonally
Screen type TFT LCD, white LED backlight, viewing cone of 60 degree and optimal viewing distance of 1 meter
Resolution 320 × 480 pixels
Alarms
Categories Patient status and system status
Priorities Low, medium, and high
Notification Audible and visual
Setting Default, institutional, and last setting
Alarm volume level 45–80 db
Alarm system delay <10s

Range and accuracy

Measurement ranges
SpO2 saturation range 1% to 100%
Pulse rate range 20 to 250 beats per minute (bpm)
Perfusion range 0.03% to 20%
Display sweep speed 6.25 mm/sec

Measurement accuracy

Saturation
Adult 70% to 100% ±2 digits
Adult and neonate low sat 60 to 80% ±3 digits
Neonate 70 to 100% ±2 digits
Low perfusion 70 to 100% ±2 digits
Adult and neonate with motion 70 to 100% ±3 digits
Pulse rate
Adult and neonate 20 to 250 bpm ±3 digits
Low perfusion 20 to 250 bpm ±3 digits
Adult and neonate with motion 20 to 250 bpm ±5 digits

Electrical

Battery Four new lithium batteries with 3,000 mAh will typically provide 20 hours of monitoring with no external communication, no audible alarm sound, and at an ambient temperature of 25 C.
Type Lithium AA
Voltage 1.5V × 4

Equipment classifications

Type of protection against electric shock Class I (internally powered)
Degree of protection against electric shock Type BF — applied part
Mode of operation Continuous
Electromagnetic compatibility IEC 60601-1-2:2007
IEC 60601-1-2:2014
Ingress protection IP22: Protected against foreign objects and moisture
Degree of safety Not suitable for use in the presence of flammable anesthetics

Ordering information

Item number Description Units per box
PM10N Patient monitoring system 1
PM10N-NA Patient monitoring kit with adult sensor 1
PM10N-PDN Patient monitoring kit with adult and pediatric sensor 1
PM10N-VSAT Patient monitoring kit with veterinary sensor 1
PMAC10N-P Protective cover — pink 1
PMAC10N-B Protective cover — blue 1
PMAC10N-N Protective cover — navy 1
PMAC10N-G Protective cover — green 1
PMAC10N-T Transport protective cover 1
PMAC10N-CC Carrying case 1

The Nellcor™ pulse oximetry monitoring system should not be used as the sole basis for diagnosis or therapy and is intended only as an adjunct in patient assessment.