Item number | Description | Units per box |
---|---|---|
PM10N | Patient monitoring system | 1 |
PM10N-NA | Patient monitoring kit with adult sensor | 1 |
PM10N-PDN | Patient monitoring kit with adult and pediatric sensor | 1 |
PM10N-VSAT | Patient monitoring kit with veterinary sensor | 1 |
PMAC10N-P | Protective cover — pink | 1 |
PMAC10N-B | Protective cover — blue | 1 |
PMAC10N-N | Protective cover — navy | 1 |
PMAC10N-G | Protective cover — green | 1 |
PMAC10N-T | Transport protective cover | 1 |
PMAC10N-CC | Carrying case | 1 |
Portable patient monitoring
Nellcor™ portable SpO₂ patient monitor
<p>The Nellcor™ portable SpO₂ patient monitor is a convenient, handheld pulse oximeter ideal for spot-checks or continuous monitoring.</p>
Features
- Small, lightweight, ergonomic handheld monitor for continuous and spot-check pulse oximetry monitoring
- Supports a robust data set, including real-time SpO2 and pulse rate values, SatSeconds alarm management, pleth waveform, blip bar, and tabular trend data
- Supports sleep study mode, which enables the clinician to dim the LCD display and silence alarms to prevent disruption of the patients’ sleep
- Compatible with the complete line of Nellcor™ single-patient use, reusable and specialty sensors with OxiMax™ technology
- 80 hours of data storage capability
- Supports data export to an external personal computer for data analysis and printing functions
- Supports standard and home-care modes for use in hospital, hospital-type facilities, transport, mobile environments, and home-care environments
Note: Transport and mobile environments are defined as intra hospital transport or transport between facilities or between a facility and a home environment
Clinical applications
Providing accurate, reliable SpO2 and pulse rate readings, this portable, easy-to-use device is an ideal solution for these critical screenings.
Specifications
Enclosure |
|
Weight | 274 g (0.604 lbs), including four batteries |
Dimensions | 70 mm W × 156 mm H × 32 mm D (2.76 in W × 6.14 in H × 1.26 in D) |
Display | |
Screen size | 88.9 mm (3.5 in), measured diagonally |
Screen type | TFT LCD, white LED backlight, viewing cone of 60 degree and optimal viewing distance of 1 meter |
Resolution | 320 × 480 pixels |
Alarms | |
Categories | Patient status and system status |
Priorities | Low, medium, and high |
Notification | Audible and visual |
Setting | Default, institutional, and last setting |
Alarm volume level | 45–80 db |
Alarm system delay | <10s |
Range and accuracy |
|
Measurement ranges | |
SpO2 saturation range | 1% to 100% |
Pulse rate range | 20 to 250 beats per minute (bpm) |
Perfusion range | 0.03% to 20% |
Display sweep speed | 6.25 mm/sec |
Measurement accuracy |
|
Saturation | |
Adult | 70% to 100% ±2 digits |
Adult and neonate low sat | 60 to 80% ±3 digits |
Neonate | 70 to 100% ±2 digits |
Low perfusion | 70 to 100% ±2 digits |
Adult and neonate with motion | 70 to 100% ±3 digits |
Pulse rate | |
Adult and neonate | 20 to 250 bpm ±3 digits |
Low perfusion | 20 to 250 bpm ±3 digits |
Adult and neonate with motion | 20 to 250 bpm ±5 digits |
Electrical |
|
Battery | Four new lithium batteries with 3,000 mAh will typically provide 20 hours of monitoring with no external communication, no audible alarm sound, and at an ambient temperature of 25 C. |
Type | Lithium AA |
Voltage | 1.5V × 4 |
Equipment classifications |
|
---|---|
Type of protection against electric shock | Class I (internally powered) |
Degree of protection against electric shock | Type BF — applied part |
Mode of operation | Continuous |
Electromagnetic compatibility | IEC 60601-1-2:2007 IEC 60601-1-2:2014 |
Ingress protection | IP22: Protected against foreign objects and moisture |
Degree of safety | Not suitable for use in the presence of flammable anesthetics |
Ordering information
Related links
The Nellcor™ pulse oximetry monitoring system should not be used as the sole basis for diagnosis or therapy and is intended only as an adjunct in patient assessment.