Your browser is out of date

With an updated browser, you will have a better Medtronic website experience.

×

Skip to main content

Check out
Bench Test data

We assessed coil’s pliability and conducted Interventional device test to quantitatively detect exertion forces on the aneurysm wall during controlled progressive coil deployment. 

Watch to learn more

0:00 / 0:00


 

Explore the Axium™ Detachable Coils family 

Axium™ Prime FC: stable &
conformable framing coils

Axium™ Prime SS: soft filling
coils (3D & helical)

Axium™ Prime ES: soft finishing
coils (3D & helical)

Axium™ Micro FX:
fibered coils (3D & helical)

Axium™ Bare: conformable &
utility coils (3D & helical)

Assisted delivery of Axium™ coils

AxiumTM family

99.7% PROVEN 
reliability of the
detachment system

(in a study including 205 Axium™ coils*)

Better trackability
through tortuous
anatomy

Tapered pusher with Contour® technology at the distal segment of pusher enhances trackability through tortuous anatomy3

Minimal
microcatheter
kick-back

Ball & socket design allows for 360° axial rotation4 which facilitates coil delivery with less microcatheter kick-back5

Progressive & Optimized Packing Volume

The entire Axium™ family features a progressive outer diameter matched to the implant size optimizes stability and volume throughout the entire coil line.

Progressive coil design: less coils, more volume

coil design

AxiumTM Prime Frame

Axium™ Prime Frame (FC)

Axium™ Prime Frame are coils with a unique loop configuration, hybrid of open & closed loops.

Versatility to conform to different aneurysm morphologies with the flexible OMEGA open loops6.

Stability enhanced with OMEGA closed-loop patterns for the first and last loops7.

Consistent and dense neck coverage as the successive loops change directions8.

axium prime frame
0:00 / 0:00

Characteristics

Coil O.D. (in)

Diameter (mm)

0.0115

3.0 - 3.5

0.0125

4.0 - 6.0

0.0135

7.0 - 10.0

0.0145

12.0 - 25.0

Coil O.D. (in)

Length (cm)

0.0115

6 - 10

0.0125

6.0 - 25.0

0.0135

12.0 - 40.0

0.0145

30.0 - 50.0

Coil O.D. (in)

Vol. (mm3)

0.0115

1.18 - 5.9

0.0125

0.59 - 5.9

0.0135

4.02 - 13.4

0.0145

4.02 - 13.4

Axium™ Prime Super Soft and Extra Soft

prime ss

 

Axium™ Prime Extra Soft and Super Soft are one of the softest coils on the market. This softness is achieved with a thinner filar and thinner stretch resistant stranders.

What are the key benefits of Axium™ Prime ES and SS coils?
  • Conformability and random loops
  • Optimal softness for filling and finishing

Characteristics

Columns
      ES-3D ES-Helix SS-3D SS-Helix
    Diameter (mm) 1 - 3.5 1 - 3 4 - 6 4 - 6
    Length (cm) 2 - 10 1 - 10 6 - 20 6 - 20
    Coil O.D. (in) 0.0108 0.0108 0.0115 0.0115
    Vol. (mm³) 1.18 - 5.9 0.59 - 5.9 4.02 - 13.4 4.02 - 13.4

    Axium™ MicroFX™

     

    Axium™ MicroFX™ coils with LatticeFX™ technology have overlapped individual nylon or PGLA microfilaments between coil loops, forming an intra-aneurysmal surface area for stable scar tissue formation and high thrombogenicity.

    These coils provide the acute performance of an Axium™ Bare coil with the added benefit of LatticeFX™ technology.

    • Flow reduction across the neck
    • Increased thrombogenicity
    • 33% less flow through MicroFX™ coils vs. standard Axium™ coils10

    axium fx

    For various situations

    Axium™ MicroFX™ coils are particularly indicated for aneurysms that have a high likelihood of rupture or are already ruptured given their high thrombogenicity and filling capacity. They should also be considered for vessel take down and fistulae as the increased filling capacity will likely reduce the procedure time and cost. 

    Finally, they can assist a flow diversion procedure where a quick thrombus formation inside the aneurysm is needed. 

    Characteristics

    Coil O.D. (in)

    PGLA 3D

    0.0115 - 0.0145

    PGLA Helix

    0.0115 - 0.0135

    Nylon Helix

    0.0115 - 0.0125

    Diameter (mm)

    PGLA 3D

    2 - 18

    PGLA Helix

    2 - 10

    Nylon Helix

    2 - 4

    Length (cm)

    PGLA 3D

    2 - 40

    PGLA Helix

    1 - 30

    Nylon Helix

    1 - 10

    Vol. (mm3)

    PGLA 3D

    1.34 - 42.61

    PGLA Helix

    0.67 - 27.7

    Nylon Helix

    0.67 - 7.92

    Axium™ Bare (QC)

    axium bare

     

    Axium™ Bare coils pioneered the Axium™ family’s benefits of high volumetric filling, progressive coil diameter as well as a highly reliable smooth and instant detachment system. The system offers the optimal balance with the softness of a “10” coil and the stability/volume of an “18” coil.

    Since 2007, several product enhancements have been made to optimize the design of Axium™ Bare (QC) coils. Their unique combination of softness and stability is obtained by balancing the size of the primary wire and the outer diameter.

    The 3D design has a large loop diameter to sphere diameter ratio and infinite, random break points.

    For various situations

    In sizes 1-6mm, QC coils have a larger OD than our Axium™ Prime line making it an ideal choice to frame and fill smaller a-morphic aneurysm where a robust coil is required and less softness is desired.

    • Enhanced conformability and reduced compartmentalization
    • Deflected into aneurysm spaces more effectively for better packing
    • Reduced “tumbling”
    • Facilitated repositioning
    • Small first loop (~70% of nominal loop diameter) anchors coil in aneurysm

    Characteristics

    Coil O.D. (in)

    3D

    0.0115 - 0.0145

    Helix

    0.0115 - 0.0145

    Diameter (mm)

    3D

    2 - 25

    Helix

    1.5 - 20

    Length (cm)

    3D

    2 - 50

    Helix

    1 - 50

    Vol. (mm3)

    3D

    1.34 - 53.27

    Helix

    0.67 - 53.27

    Delivering Axium™ Coils

    Balloon-Assisted coiling

    With our neuro-balloons, apply supple conformity to complex structures. Our neuro-balloons are ideal for distal access, and unique and irregular anatomies, such as bifurcation geometries. They provide stability in large and small vessels. Discover dependable coverage specifically designed to create stable occlusion across the challenging diameters encountered in large and small vessels.

    Our HyperForm™ occlusion balloons are created with extra softness to conform to asymmetrical, irregular and complex vasculature geometries.

    Our HyperGlide™ occlusion balloons are oblong compliant balloons designed for sidewall aneurysms and reliable occlusions.

    balloon coil

    Indications

    The Occlusion Balloon Catheters are indicated for use in blood vessels of the peripheral and neuro vasculature where temporary occlusion is desired. These catheters offer a vessel selective technique of temporary vascular occlusion, which is useful in selectively stopping or controlling blood flow; the Occlusion Balloon Catheters may also be used in balloon-assisted embolization of intracranial aneurysms and treatment of vasospasm.

    See the device manual for detailed information regarding the instructions for use, indications, contraindications, warnings, precautions, and potential adverse events.

    Characteristics

    Usable Length (cm): 150

    Proximal Outer Diameter (F)

    Hyperform™

    2.8

    HyperGlide™

    2.8

    Distal Outer Diameter (F)

    Hyperform™

    2.2 - 3.0

    HyperGlide™

    2.2

    Balloon Diameter (mm)

    Hyperform™

    3.0 - 7.0

    HyperGlide™

    3.0 - 5.0

    Balloon Length (mm)

    Hyperform™

    7.0 - 15.0

    HyperGlide™

    10.0 - 30.0

    Tip Length (mm)

    Hyperform™

    2.0

    HyperGlide™

    4.0

    stent coil

    Stent-Assisted coiling

    Solitaire™ AB device is a self-expanding stent with a unique design to bridge the neck of aneurysms that can be completely retrieved, even when fully deployed for unmatched procedural control.

    The Solitaire™ AB neurovascular remodeling device, with the unique Parametric™ technology, is the adequate solution for assisting coiling procedures by acting as a physical barrier in wide-neck and blister aneurysms to trap coils and prevent coil herniation into the parent artery.

    Indications

    The Solitaire™ AB Neurovascular Remodeling Device is designed for use as an adjunctive device in the treatment of intracranial aneurysms.

    See the device manual for detailed information regarding the instructions for use, indications, contraindications, warnings, precautions, and potential adverse events.

    Available CFNs

    CFN

    SAB-3-20

    Vessel Diameter Range (mm)

    2.0 - 3.0

    Stent Diameter (mm)

    3.0

    Min. Microcatheter Inner Diameter (in)

    0.021

    Usable Length (mm)

    20

    Distal Markers

    3

    Proximal Markers

    1

    CFN

    SAB-3-30

    Vessel Diameter Range (mm)

    2.2 - 3.0

    Stent Diameter (mm)

    3.0

    Min. Microcatheter Inner Diameter (in)

    0.021

    Usable Length (mm)

    30

    Distal Markers

    3

    Proximal Markers

    1

    CFN

    SAB-4-15

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.021

    Usable Length (mm)

    15

    Distal Markers

    3

    Proximal Markers

    1

    CFN

    SAB-4-20

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.021

    Usable Length (mm)

    20

    Distal Markers

    3

    Proximal Markers

    1

    CFN

    SAB-4-30

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.021

    Usable Length (mm)

    30

    Distal Markers

    3

    Proximal Markers

    1

    CFN

    SAB-4-40

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.021

    Usable Length (mm)

    40

    Distal Markers

    3

    Proximal Markers

    1

    CFN

    SAB-5-20

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.027

    Usable Length (mm)

    20

    Distal Markers

    4

    Proximal Markers

    1

    CFN

    SAB-5-30

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.027

    Usable Length (mm)

    30

    Distal Markers

    4

    Proximal Markers

    1

    CFN

    SAB-5-40

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.027

    Usable Length (mm)

    40

    Distal Markers

    4

    Proximal Markers

    1

    CFN

    SAB-6-20

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.027

    Usable Length (mm)

    20

    Distal Markers

    4

    Proximal Markers

    1

    CFN

    SAB-6-30

    Vessel Diameter Range (mm)

    3.0 - 4.0

    Stent Diameter (mm)

    4

    Min. Microcatheter Inner Diameter (in)

    0.027

    Usable Length (mm)

    30

    Distal Markers

    4

    Proximal Markers

    1

    The Solitaire™ AB detachment system is a battery operated device working with detachment cables designed to initiate and control the detachment of the Solitaire™ AB neurovascular remodeling device.

    CFN

    NDS-2X

    Product

    Solitaire™ Detachment Box

    CFN

    CSS-2.75-1X

    Product

    Cable Set, Length: 2.75M, Quantity 1

    CFN

    CSS-2.75-5X

    Product

    Cable Set, Length: 2.75M, Quantity 5

    Specifications

    Axium™ and Axium™ Prime detachable coil systems are indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae. 

    Axium™ and Axium™ MicroFX™ detachable coil systems are indicated for the endovascular embolization of intracranial aneurysms and other neurovascular abnormalities, such as arteriovenous malformations and arteriovenous fistulae.

    Any questions?

    Do you need support for procedures? Or information on our products and solutions?

    Our team is happy to help answer any questions you may have.

    Looking for
    more in
    Neurovascular?

    Go to Neurovascular


    This material is for Healthcare Professionals in countries with applicable health authority product registrations.

    Important: Always refer to the Instructions For Use (IFU) packaged with the product/e-IFU for complete instructions, indications, contraindications, warnings, and precautions. 

    1

    Data on file. TR-NV12481A /

    2

    Data on file TR-07-053A & TR-09-218A

    3

    TR-NV12481A; TR10-126A

    4

    TR-NV12481A; FD3117A; FD3118A. Based on bench test comparing axial rotation of Axium™ Prime coil versus Target® coils.1 unit of each coil were tested.
    Bench testing is not indicative of clinical performance

    5

    Data on file. TR-NV12481A /

    6

    Data on file Top Level Drawing 92566A. Ref: MP1936B & RE00046057 Data on file. TR-NV13289A & Addendum

    7

    Ref: Data on file Top Level Drawing & RE00046057

    8

    Data on file Top Level Drawing 92566A. MP1936B & RE00046057

    9

    Fargen KM, Blackburn S, Carpenter JS, et al. J NeuroIntervent Surg Published Online First: doi:10.1136/neurintsurg-2013-010887

    10

    Girdhar G, et al, In-vitro thrombogenicity assessment of polymer filament modified and native platinumembolic coils, J Neurol Sci (2014)

    * RE00032057; TR-NV13164; TR09-218A. In a study including 205 Axium™ coils, 100% detached on the first attempt. The initial detachment reliability is at least 98.5% reliable with 95% confidence