비디오를 재생합니다. - REVEAL LINQ

REVEAL LINQTM

Insertable Cardiac Monitor

MRI CONDITIONS FOR USE

The Reveal LINQ device is MR Conditional and, as such, is designed to allow patients to be safely scanned by an MRI machine when used according to specified MRI conditions for use. The MR Conditional symbol shown below is used to indicate the conditional safety of devices and components in the MR environment.

Conditional symbol

MR

A patient with a Reveal LINQ device can be safely scanned in an MR system that meets the following conditions.

Failure to follow these conditions for use may result in a hazard to the patient during an MRI scan:

  • Horizontal cylindrical bore magnet, clinical MRI systems with a static magnetic field of 1.5 Tesla (T) or 3.0 T must be used.
  • Hydrogen proton MRI equipment must be used.
  • Maximum spatial gradient of the static magnetic field specification must be ≤25 T/m (2500 gauss/cm).
  • Whole body gradient systems with gradient slew rate specification must be ≤200 T/m/s per axis.
  • The Whole Body Specific Absorption Rate (WB-SAR) as reported by the MRI equipment must be ≤4.0 W/kg; the head SAR as reported by the MRI equipment must be ≤3.2 W/kg.
  • Do not use local transmit coils on the chest, trunk, or shoulder region.
  • There are no restrictions on the placement of receive-only coils, and there are no restrictions on the use of local transmit or receive coils for imaging of the head or extremities.