EffectivCRT Diagnostic Cardiac Device Features

Overview

A high percentage of BiV pacing is associated with improved CRT response.1 However, the presence of a pacing stimulus does not imply full capture.2,3 Traditional pacing counters only report the presence of a pacing stimulus, potentially leading to over-reporting of effective pacing.2,3

You can now verify effective pacing with the EffectivCRT Diagnostic

The EffectivCRT™ Diagnostic uses a unipolar electrogram to evaluate morphology, looking for a negative deflection which implies when a paced beat is effective.4

  • This EGM strip shows 6 consecutive paced beats
  • Traditional pacing counters would report this as 100% V. pacing
  • With the EffectivCRT Diagnostic, it would report 100% V pacing but 50% is effective pacing
    • The first 3 beats are ineffective (positive deflection implies tissue is paced and not captured)
    • The last 3 beats are effective (negative deflection implies tissue is paced and captured)
EGM strip shows 6 consecutive paced beats. The first 3 are ineffective. The second 3 are effective.

In 18% of patients, device-reported % V. Pacing overestimated effective CRT by 3%. 3 out of 57 patients had virtually no effective pacing, whereas the % V pacing was > 90%, as demonstrated in the graph below.5

Before the
EffectivCRT™ diagnostic:

Only see the quantity of pacing5

before-effectivcrt-diagnostic graphic

With the
EffectivCRT™ diagnostic:

Also see the quality of pacing5

with-effectivcrt-diagnostic graph
You can use the data provided by the EffectivCRT Diagnostic to identify ineffective pacing and will find it in these places:
  • Cardiac Compass™ Trends
  • Quick Look™ II Screen
  • Rate Histogram Report
  • EffectivCRT Episodes
1

Hayes DL, et al. Heart Rhythm. 2011;8:1469-1475.

2

Daubert JC, et al. Heart Rhythm. 2012;9:1524–1576.

3

Brignole M, et al. Eur Heart J. 2013;34:2281–2329.

4

Ghosh S, et al. Europace. 2015;17:1555-1562.

5

Hernandez, et al. Presented at CardioStim 2016 (Abstract 99-06).