We work closely with our OEM partners to design integrated monitoring solutions knowing that the best possible outcomes happen when all parts of a system work well together.

From our sensors to the front-end electronics and the signal processing, Nellcor™ pulse oximetry technology is engineered to provide accurate and reliable, real-time SpO2 readings, in a variety of patient conditions such as neonates and adults in motion.([FOOTNOTE=Vesoulis Z, Lust C, Liao S, Trivedi S, Mathur A.  Early hyperoxia burden detected by cerebral near-infrared spectroscopy is superior to pulse oximetry for prediction of severe retinopathy of prematurity. J Perinatol. 2016. 36:966.],[ANCHOR=],[LINK=]) ,([FOOTNOTE=Medtronic Clinical Investigations:
▪ Doc no. 10101937, NCT01720355 at ClinicalTrials.gov
▪ Doc no. 10047614 (Motion and Low Perfusion with OxiMax N-595
▪ Doc no. 10096850 (Low Sat Accuracy Connery 3.2 boards with N600x)
▪ Doc no. 10099560 (Max Sensors- COVMOPR0384 Motion LAMP-C)
▪ Doc no. 10034561
▪ Invasive Controlled Hypoxia Study Summary Report – MX60: Validate Across Abbreviated Sensor Line vs. CO-OX (attachment R-J1)
▪ Invasive Controlled Hypoxia Study Summary Report – Low Sat BC: NELL-1 (MP100), NELL-3 (MP506) w/ MAX-A, MAXN, SC-A, MAX-FAST; N600 w/ MAX-A vs. CO-OX (attachment R-J2)
▪ Clinical Summary Report, Jupiter, N600x (attachment R-J3) (Doc no. 10011350)
▪ Doc no. 1002061, OxiMax Oxygen Saturation Accuracy Specifications],[ANCHOR=],[LINK=])

Features

Offers numeric arterial oxygen saturation and pulse rate measurement values, and plethysmogram waveform on Philips IntelliVue™* patient monitors.

Compatible with:
  • Philips IntelliVue™* MP40, MP50, MP60, MP70, MP90 monitors
  • Philips IntelliVue™* MX500, MX550, MX700, MX750, MX800, MX850 monitors
Order Information
Order Code Description Unit of Measure Quantity
10076644 Nellcor™ SpO2 Module with Interface Cable, for use with Philips Intellivue™* Monitor Box 1
10083058 Nellcor™ SpO2 Module without Interface Cable, for use with Philips Intellivue™* Monitor Box 1
Physical
Dimensions: 3.80 in. x 3.67 in. x 1.39 in (9.65 cm x 9.32 cm x 3.53 cm)
Weight: 7.2 oz  (210 g)
Safety
Complies with IEC 60601-1, IEC 60601-1-2, CSA C22.2#601.1-M90AMD2, ISO 9919, IEC 80601-2-61
Type CF applied part, defibrillator proof
U.S. Federal law restricts this device to sale by or on the order of a physician.
SpO2 Sensors
Wavelength range: 500 to 1,000 nm
Emitted light energy: ≤ 15 m W
Environmental
Module
Operating temperature: 0 to 45 C (32 to 113 F)
Storage temperature: – 40 to 70 C (– 40 to 158 F)
Operating humidity: 95% RH max. at 40 C (104 F)
Storage humidity: 95% RH max. at 65 C (150 F)
Operating altitude: Up to 4,600 m (15,000 ft)
Storage altitude: Up to 15,300 m (50,000 ft)
Compatibility
Philips IntelliVue™* MP40, MP50, MP60, MP70, MP80, MP90 monitors
Philips IntelliVue™* MX800 monitors

Order Information

Technology