Ventilator Settings |
---|
Ideal body weight (IBW) |
0.3 to 7.0 kg with NeoMode 2.0 (0.66 to 15 lbs), 7.0 to 24 kg (15 to 53 lbs), 25 to 150 kg (55 to 330 lbs) |
Modes |
Assist/Control (A/C), SYNCHRONISED INTERMITTENT MANDATORY VENTILATION intermittent mandatory ventilation (SIMV), SPONTANEOUS (SPONT), Bi-Level |
Software Options |
NeoMode 2.0, Leak Compensation, Tube Compensation, Bi-Level, Volume Ventilation Plus (Volume Control Plus and Volume Support), Proportional Assist™* Ventilation Plus (PAV+™), Respiratory Mechanics and Trending |
Mandatory breath types |
Volume Control (VC), Pressure Control (PC) or Volume Control Plus |
Spontaneous breath types |
Pressure Supported (PS), Volume Supported (VS), Proportional Assist (PA) |
Vent type |
Invasive or Noninvasive |
Pressure support (PSUPP) |
0 to 70 cm H2O |
Rise time % |
1% to 100% |
Expiratory sensitivity (ESENS) |
1% to 80%; 1 L/min to 10 L/min with PAV™*+ |
Tidal volume (VT) |
25 to 2,500 mL, 2 to 315 mL with NeoMode
|
Respiratory rate (f) |
1.0 to 100 /min, 1 to 150 /min with NeoMode |
Peak inspiratory flow (VMAX) |
3 to 150 L/min for IBW > 24 kg; 3 to 60 L/min for IBW ≤ 24 kg, 1 to 30 L/min with NeoMode |
Flow pattern |
Square or descending ramp |
Plateau time (TPL) |
0.0 to 2.0 secondsPlateau time (TPL): 0.0 to 2.0 seconds |
Inspiratory pressure (PI) |
5 to 90 cm H2O |
Constant during rate change |
Inspiratory time (TI), I:E ratio or expiratory time |
Inspiratory time (TI) |
0.2 to 8.0 seconds |
I:E ratio |
≤ 1:299-4.00:1 |
Expiratory time (TE) |
TE ≥ 0.2 second |
Trigger type |
Pressure (PTRIG) or flow (VTRIG Flow-by flow triggering) |
ressure sensitivity (PSENS) |
0.1 to 20 cm H2O below PEEP |
Flow sensitivity (VSENS) |
0.2 to 20 L/min, 0.1 to 10 L/min with NeoMode |
O2% |
21% to 100% |
PEEP |
0 to 45 cm H2O |
Apnea ventilation |
Apnea mandatory type-volume control (VC) or pressure control (PC)
|
Apnea flow pattern |
Square or descending ramp |
Apnea peak flow (VMAX) |
≥ 1.0 L/min to ≤ 30 L/min, ≥ 3.0 L/min to ≤ 60 L/min, ≥ 3.0 L/min to ≤ 150 L/min |
Apnea tidal volume (VT) |
3 mL to 315 mL for neonatal; 25 mL to 2500 mL for ped/adult |
Apnea inspiratory pressure (PI) |
5 to 90 cm H2O |
Apnea inspiratory time (TI) |
0.2 to 8.0 seconds |
Apnea interval (TA) |
10 to 60 seconds |
Apnea respiratory rate (f) |
2.0 to 40 1/min |
Apnea O2% |
21% to 100% |
Apnea expiratory time (TE) |
≥ 0.2 seconds |
Disconnect sensitivity (DSENS) |
20% to 95% or OFF in NIV |
Humidification type |
Heat-moisture exchanger (HME), nonheated expiratory tube or heated expiratory tube |
Humidification volume |
100 to 1,000 mL |
Patient circuit type |
Pediatric, adult or neonate with NeoMode option |