Overview

Must-have Features in a Convenient, Portable Design

The Nellcor™ portable SpO2 patient monitoring system, PM10N, is a convenient, handheld monitor that is ideal for spot-checks or continuous monitoring in various healthcare and home use settings. Its ergonomic shape and simple design make it intuitive to use and simple to operate. Wherever you need accurate, reliable SpO2 measurements, count on the Nellcor™ portable SpO2 pulse oximeter to meet the challenge.

The monitoring system includes a vivid 3-inch color LCD screen, and connectivity to analytic tools and patient management systems. It is compatible with the entire line of Nellcor™ sensors with OxiMax™ technology and offers a robust monitoring feature set including SpO2, pulse rate, SatSeconds alarm management, pleth waveform, blip bar and tabular trend information.

The monitor incorporates Nellcor™ digital signal processing technology to deliver accurate, reliable SpO2 and pulse rate values even during low perfusion and signal interference, including patient motion.

Product Features

  • Small, lightweight, ergonomic handheld monitor for continuous and spot-check pulse oximetry monitoring
  • Supports a robust data set, including real-time SpO2 and PR values, SatSeconds alarm management, pleth waveform, blip bar and tabular trend data
  • Supports standard and home-care modes for use in hospital, hospital-type facilities, transport, mobile environments and home-care environments
  • Supports Sleep Study Mode, which enables the clinician to dim the LCD display and silence alarms to prevent disruption of patients’ sleep
  • Nellcor™ advanced signal processing technology delivers reliable operation even in difficult monitoring conditions
  • LoSat expanded accuracy range (60% to 100% SpO2) when used with Nellcor™ adhesive sensors with OxiMax™ technology
  • Compatible with the complete line of Nellcor™ single-patient-use, reusable and specialty sensors with OxiMax™ technology
  • 80 hours of data storage capability
  • Supports data export to an external personal computer for data analysis and printing functions

How to Order

For additional information or to place an order, call Customer Service at +44 (0) 19 2320 2504.

Order Information
Order Code Description Unit of Measure Quantity
PM10N Nellcor™ Portable SpO2 Patient Monitoring System Box 1
PMAC10N-P Portable SpO2 Protective Cover (Pink) Box 1
PMAC10N-B Portable SpO2 Protective Cover (Blue) Box 1
PMAC10N-N Portable SpO2 Protective Cover (Navy) Box 1
PMAC10N-G Portable SpO2 Protective Cover (Green) Box 1
PMAC10N-T Portable SpO2 Transport Protective Cover Box 1
PMAC10N-CC Portable SpO2 Carrying Case  Box 1
Specifications
Enclosure
Weight 274 g (0.604 lbs), including four batteries
Dimensions 70 mm W x 156 mm H x 32 mm D
  (2.76 in W x 6.14 in H x 1.26 in D)
Display  
Screen size 88.9 mm (3.5 in), measured diagonally
Screen type TFT LCD, white LED backlight, viewing cone of 60º and optimal viewing distance of 1 meter
Resolution 320 x 480 pixels

Alarms  
Categories Patient status and system status
Priorities Low, medium and high
Notification Audible and visual
Setting Default, institutional and last setting
Alarm volume level 49 to 89 dB
Alarm system delay <10s
Range and Accuracy
Range Type Range Values
Measurement Ranges
 
SpO2 saturation range 1% to 100%
Pulse rate range 20 to 250 beats per minute (bpm)
Perfusion range 0.03% to 20%
Display sweep speed 6.25 mm/sec
Measurement Accuracy
Saturation  
Adult 70% to 100% ±2 digits
Adult and neonate low sat 60 to 80% ±3 digits
Neonate 70 to 100% ±2 digits
Low perfusion 70 to 100% ±2 digits
Adult and neonate with motion 70 to 100% ±3 digits
Pulse rate  
Adult and neonate 20 to 250 bpm ±3 digits
Low perfusion 20 to 250 bpm ±3 digits
Adult and neonate with motion 20 to 250 bpm ±5 digits
Electrical
Battery Four new lithium batteries with 3,000 mAh will typically provide 20 hours of monitoring with no external communication, no audible alarm sound and at an ambient temperature of 25° C.
Type Lithium AA
Voltage 1.5V x 4
Environmental
  Transport and Storage Operating Conditions
Temperature -20 ºC to 70 ºC, 5 ºC to 40 ºC
  (-4 ºF to 158 ºF) (41 ºF to 104 ºF)
Altitude -390 to 5,574 m, -390 to 5,574 m
  (-1,280 to 18,288 ft.) (-1,280 to 18,288 ft.)
Relative humidity 15% to 95% non-condensing
Trends
Types Tabular
Memory Saves total 80 hours data events. Saves date and time, alarm conditions, pulse rate and SpO2 measurements
Tabular format One table for all parameters
Standards Compliance  
IEC 60601-1:2005+A1:2012, EN 60601-1:2006/AC:2010
IEC 60601-1:1998 + A1:1991 +A2:1995, EN 60601-1:1990 +A11:1993 +A12:1993 +A13:1996
IEC 60601-1-2:2007, EN60601-1-2:2007
IEC 60601-1-6:2010, EN 60601-1-6:2010 +A1:2013
IEC 60601-1-8:2006, EN 60601-1-8:2006 +A1:2012
IEC 60601-1-11:2010, EN 60601-1-11:2010
ISO 9919:2005, EN ISO 9919:2009
ISO 80601-2-61:2011, EN ISO 80601-2-61:2011
CAN/CSA C22.2 No. 601.1 M90
UL 60601-1: 1st edition
802.11 B/G/N WLAN connectivity
Equipment Classifications
Type of protection against electric shock Class I (internally powered)
Degree of protection against electric shock Type BF - applied part
Mode of operation Continuous
Electromagnetic compatibility IEC 60601-1-2:2007
Ingress protection IP22: Protected against foreign objects and moisture
Degree of safety Not suitable for use in the presence of flammable anesthetics