Compact size, low volume and high efficiency mechanical filter.
For single patient use on anesthetized patients and respiratory care patients who require a breathing circuit. The product is intended to be replaced at least once every 24 hours.
Order Code | Description | Size | Connection | Unit of Measure | Quantity |
---|---|---|---|---|---|
351/5979 | DAR™ Mechanical Filter Small | Vт 150 - 1200 ml | ISO 22M/15F - 22F/15M | Box | 25 |
351/5979TC | DAR™ Mechanical Filter Small with tethered cap | Vт 150 - 1200 ml | ISO 22M/15F - 22F/15M | Box | 25 |
Resistance to flow (ISO 9360) | |
---|---|
30 L/min | 1.2 cm H2O |
60 L/min | 2.7 cm H2O |
Bacterial filtration efficiency | 99.99999% |
Viral filtration efficiency | 99.997% |
Dead space (connectors included-ISO 9360) | 42 ml |
Weight | 24 g |
Type of filtration | Mechanical |