Overview

Effective moisture control reduces the risk of colonization in breathing tubes and reduces sampling line blockage.

Features and benefits

  • Combined filter / HME sized for infants reduces set up time, and reduces dead space and weight while increasing efficiency
  • Hygroscopic membrane traps patient's moisture providing effective airway humidification
  • Low resistance to airflow reduces work of breathing
  • Effective moisture control reduces the risk of colonization in breathing tubes and reduces sampling line blockage
  • End tidal CO2 sampling port offers convenient access to airway gases
  • ISO standard 15 mm fitting connects with breathing system
  • Supplied clean and individually packaged

Indications for usage

  • For infant (recommended tidal volume 20 - 70 ml), single patient use on anesthetized patients and respiratory care patients who require a breathing circuit. The product is intended to be replaced at least once every 24 hours.

How to Order

For additional information or to place an order, call Customer Service at +44 (0) 19 2320 2504.

Order Information
Order Code Description Size Connection Unit of Measure Quantity
355/5427 DAR™ Pediatric - Neonatal Electrostatic Filter HME Small Vт 30 - 100 ml ISO 15F/15M Box 25
355/5427TC DAR™ Pediatric - Neonatal Electrostatic Filter HME Small with tethered cap Vт 30 - 100 ml ISO 15F/15M Box 25
Specifications
Recommended tidal volume 30 - 100 ml
Moisture output and temperature (ISO 9360)
Vt 50 ml 2 hours - 30.0 mg H2O/L at 30.2°C
Resistance to flow (ISO 9360) 0.6 cm H2O at 5 L/min
Bacterial filtration efficiency 99.999%
Viral filtration efficiency > 99.99%
Dead space (connectors included - ISO 9360) 10 ml
Weight 9 g
Type of filtration Electrostatic

How to Order

For additional information or to place an order, call Customer Service at +44 (0) 19 2320 2504.

Order Information
Order Code Description Size Connection Unit of Measure Quantity
355/5427 DAR™ Pediatric - Neonatal Electrostatic Filter HME Small Vт 30 - 100 ml ISO 15F/15M Box 25
355/5427TC DAR™ Pediatric - Neonatal Electrostatic Filter HME Small with tethered cap Vт 30 - 100 ml ISO 15F/15M Box 25
Specifications
Recommended tidal volume 30 - 100 ml
Moisture output and temperature (ISO 9360)
Vt 50 ml 2 hours - 30.0 mg H2O/L at 30.2°C
Resistance to flow (ISO 9360) 0.6 cm H2O at 5 L/min
Bacterial filtration efficiency 99.999%
Viral filtration efficiency > 99.99%
Dead space (connectors included - ISO 9360) 10 ml
Weight 9 g
Type of filtration Electrostatic