The ClosureFast™ procedure, with its patented overlap design, is the only RFA procedure with published long-term clinical data demonstrating safety and efficacy, with a 91.9% closure rate at 5 years.1,2
Connect with usGap-free thermal ablation
The ClosureFast™ RFA system delivers uniform heat for lasting vein closure.1
Catheter inserted into vein
Controlled heat collapses vein
Catheter withdrawn, closing vein
ClosureFast™ endovenous radiofrequency ablation procedure
The ClosureFast™ procedure features a patented overlap of 0.5 cm at each thermal treatment segment, eliminating gaps between segments ― for lasting vein closure.1,2
Both 3 cm and 7 cm heating elements provide an overlap of 0.5 cm as the device is repositioned.
Note: Catheter also available with 3 cm heating element and 2.5 cm markings.
Building long-term evidence with high clinical standards
See More Clinical EvidenceThe ClosureFast™ endovenous radiofrequency ablation (RFA) catheter is intended for endovascular coagulation of blood vessels in patients with superficial vein reflux. The catheter is provided sterile, and is a single-use, disposable device. This catheter cannot be adequately cleaned and/or sterilized by the user in order to facilitate safe reuse, and is therefore intended for single use. Attempts to clean or sterilize these devices may result in a bio-incompatibility, infection, or product failure risks to the patient.
Catheter Model |
CF7-7-60 |
CF7-7-100 |
CF7-3-60 |
---|---|---|---|
Introducer Sheath |
7 F |
7 F |
7 F |
Insertable Length |
60 cm |
100 cm |
60 cm |
Heating Element Diameter |
2.3 mm |
2.3 mm |
2.3 mm |
Heating Element Length |
7 cm |
7 cm |
3 cm |
Heating Element Overlap |
0.5 cm |
0.5 cm |
0.5 cm |
Compatible Guidewire |
0.025" |
0.025" |
0.025" |
Maximum Power Setting |
40 W |
40 W |
18 W |
Default Target Temperature Setting |
120°C |
120°C |
120°C |
Software Version: RFG2 |
4.0.0 or higher |
4.0.0 or higher |
4.0.0 or higher |
Software Version: RFG3 |
1.11.0 or higher |
1.11.0 or higher |
1.11.0 or higher |
Connect with a Medtronic representative and receive updates and ongoing information.
Connect NowProebstle TM, Alm BJ, Göckeritz O, et al. Five-year results from the prospective European multicentre cohort study on radiofrequency segmental thermal ablation for incompetent great saphenous veins. Br J Surg. February 2015;102(3):212-218.
ClosureFast and ClosureFast RFS Patents. Medtronic data on file, 2022.
Almeida JI, Kaufman J, Göckeritz O, et al. Radiofrequency endovenous ClosureFAST versus laser ablation for the treatment of great saphenous reflux: A multicenter, single-blinded, randomized study (RECOVERY Study). J Vasc Interv Radiol. June 2009;20(6):752-759.
Morrison N, Gibson K, Vasquez M, et al. Five-year extension study of patients from a randomized clinical trial (VeClose) comparing cyanoacrylate closure versus radiofrequency ablation for the treatment of incompetent great saphenous veins. J Vasc Surg Venous Lymphat Disord. November 2020;8(6):978-989.
Medtronic ClosureFast Procedure Publications. Medtronic data on file, 2019.
ClosureFast Procedures. Medtronic data on file, 2019.