GENERAL DATA |
---|
DIMENSIONS |
161 x 43 x 46 (H x W x D in cm)
Base: 51 x 74 (W x D in cm) |
---|
Weight |
Approx. 90 kg (incl. all options)
|
---|
Water supply |
Pressure: 1 to 7 bar at minimum 800 mL / min at maximum 3000 mL / min
Temp.: 5 to 30°C
|
---|
Drain |
Minimum drain capacity: 800 mL / min average
Height: 50 cm maximum
Temp.: 90 ° C maximum
|
---|
Concentrate supply |
Pressure: 0 to 0.5 bar
2 central acid concentrates
|
---|
Power supply |
110 to 120 VAC ±10 %, 50 to 60 Hz ±1 Hz (≈13 A)
|
---|
Battery |
Ni-MH battery 24 V / 3200 mAh
|
---|
External connection port |
External output (staff call)
External input 1
External input 2
Nurse call switch
LAN/Network (RJ-45)
Serial interface (RS-232)
BPM start switch
USB
CF card type I
|
---|
PAS |
Input interface for external alarming device (IEC PAS 63023) to stop extracorporeal and / or fluid circuit
|
---|
Monitor |
15 inch LCD
|
---|
Hydraulic circuit |
---|
Dialysis fluid flow rate |
Setting range:
Single ETRF 300 to 800 mL / min
Double ETRF 300 to 700 mL / min
|
---|
Dialysis fluid temperature |
Setting range: 34.0 to 40.0 ° C
|
---|
Dialysis fluid conductivity |
Bicarbonate dialysis
Bicarbonate conductivity setting range: 2.3 to 7.0 mS / cm
Accuracy: ±0.1 mS / cm
Total conductivity setting range: 12.7 to 15.2 mS / cm
Accuracy: ±0.2 mS / cm
Acetate dialysis
Total conductivity setting range: 12.7 to 15.2 mS / cm
Accuracy: ±0.2 mS / cm
|
---|
Transmembrane pressure (TMP) |
Measurement range: -100 to +500 mmHg
Measurement accuracy: ±10 mmHg
|
---|
Blood leak detector |
Method: Optical
Sensitivity: 0.3 mL Blood / 1 L Dialysis fluid
(Blood: Hematocrit 32 ± 2 %; Dialysis fluid temperature: 37° C)
|
---|
Ultrafiltration |
UF rate: 0.00; 0.10 to 4.00 L / h
UF accuracy (Balance):
±30 mL / h (At dialysis fluid flow rate 300 to 500 mL / min)
±0.1 % of the dialysis fluid flow rate (At dialysis fluid flow rate 501 to 800 mL / min)
|
---|
Dialysis Dose Monitor |
Measurement principle: Absorptiometry
Applicable Treatment mode: HD, On-line HDF
Applicable Kt / V range: 0 to 3.0
Kt/V monitoring accuracy: ±0.1 (Kt / V 0 to 1)
±10 % (Kt / V 1 to 3)
Applicable URR range: 0 % to 100 %
URR monitoring accuracy: ±5 %
|
---|
Endotoxin retentive filter (ETRF) |
EF-02D
|
---|
Treatment options |
---|
Online HDF/HF |
Substitution flow setting range: 0.00; 0.10 to 18.00 L / h (Online HDF)
0.00; 0.10 to 30.00 L / h (Online HF)
Flow rate accuracy: ±10 % of set value
|
---|
Single needle treatment |
Single needle single pump treatment
Single needle double pump treatment
SN control pressure: Upper limit: +400 mmHg
Lower limit: 0 mmHg
|
---|
UF profiles |
9 programmable profiles available
|
---|
Conductivity profiles |
9 programmable profiles available
|
---|
Extracorporeal circuit |
---|
Arterial pressure monitoring |
Measurement range: -300 to +500 mmHg
Measurement accuracy: ±10 mmHg
|
---|
Venous pressure monitoring |
Measurement range: -300 to +500 mmHg
Measurement accuracy: ±10 mmHg
|
---|
Dialyser inlet blood pressure monitoring |
Measurement range: -300 to +735 mmHg
Measurement accuracy: ±10 mmHg
|
---|
Single needle pressure |
Measurement range: -200 to +600 mmHg
Measurement accuracy: ±10 mmHg
|
---|
Air detector |
Method: Ultrasonic waves
Sensitivity: 0.02 mL (normal air bubbles) (At Blood flow rate: 250 mL / min)
0.0003 mL (microbubbles: blood / air mixture)
(At Blood flow rate: 250 mL / min)
|
---|
Arterial blood pump (PUMP1) |
Setting range: 40 to 600 mL / min
Flow rate accuracy: Set value ±10 % (inlet Pressure -150 mmHg P +150 mmHg)
Set value -20 to 0 % (inlet Pressure -200 mmHg P -150 mmHg)
|
---|
Heparin pump |
Setting range: 0.0 to 9.9 mL / h
Output rate accuracy: Set value ±10 %
Syringe type:
30 mL or 20 mL,
20 mL or 10 mL (optional)
Bolus volume: 0.0 to 9.9 mL
|
---|
Venous blood pump / Substitution fluid pump (PUMP2) |
Setting range: 40 to 600 mL / min
Flow rate accuracy: Set value ±10 % (inlet Pressure -150 mmHg P +150 mmHg)
Set value -20 to 0 % (inlet Pressure -200 mmHg P -150 mmHg)
|
---|
Blood Pressure Monitor (BPM) |
Pressure display range: 10 to 300 mmHg
Pressure display accuracy: Less than ±3 mmHg
Measurement range (adults):
Systolic blood pressure (SYS): 60 to 250 mmHg
Mean arterial pressure (MAP): 45 to 235 mmHg
Diastolic blood pressure (DIA): 40 to 200 mmHg
Pulse rate: 40 to 200 beats per minute
|
---|
For iNIBP:
Pressure display range: 0 to 300 mmHg
Pressure display accuracy: Less than ±3 mmHg
Measurement range:
Adult systolic blood pressure (SYS): 40 to 280 mmHg
Mean arterial pressure (MAP): 10 to 280 mmHg
Diastolic blood pressure (DIA): 10 to 235 mmHg
Pulse rate: 30 to 200 beats per minute
|
Blood Volume Monitor (BVM) |
Measurement principle : Near-infrared reflection method
Applicable blood flow rate range: 40 to 600 mL / min
Applicable hematocrit range: 15 to 50 %
Accuracy: ±2.3 dBV % (Double needle)
|
---|
Vascular access recirculation rate measurement |
Usable treatment modes: HD, ISO-UF, OHF, OHDF
Measurement range: 0~100 %
Measurement accuracy: ±10 (recirculation ratio %)
Measurement times: Maximum 5 times
Usable blood flow rate range: 100~600 ml / min
Usable hematocrit value range: BVM measurement range 15 to 50 %
Usable blood tubing lines: NIKKISO AV18-series
|
---|
Accessories |
---|
Hook for concentrate bags |
Max. load 10 kg
|
---|
Patient card |
MIFARE Classic 4K
Capacity: 4096 bytes
|
---|
Cleaning program |
---|
Disinfection and decalcification |
Heat disinfection with 50 % Citric acid |
---|
Disinfection and degreasing |
Sodium hypochlorite solution (Maximum 10 %)
|
---|
Decalcification |
30 % Acetic acid
|
---|