Your browser is out of date

With an updated browser, you will have a better Medtronic website experience. Update my browser now.

×

Skip to main content

Healthcare Professionals

Sprinter Legend RX

Sprinter Legend RX Semicompliant Balloon Dilatation Catheter

Sprinter Legend RX Semicompliant Balloon Dilatation Catheter

With innovative crossing technologies and 57 sizes, the Sprinter® Legend RX Semicompliant Balloon Dilatation Catheter provides you with the power to cross.

How the Sprinter Legend RX Catheter Works

Percutaneous Coronary Intervention (PCI), also known as coronary angioplasty, is a therapeutic procedure to treat the narrowed or stenotic section of the coronary artery of the heart due to coronary atherosclerosis. In balloon catheter angioplasty, tightly folded and narrow balloons are passed into the narrowed artery and inflated to a fixed size using liquid. This procedure reduces the narrowing of the artery by pushing the plaque (or blockage) to the sides of the artery.1

Sprinter Legend RX Percutaneous Coronary Intervention

Stent implantation often follows balloon angioplasty. Once the stent is implanted, post stent deployment as assessed by IVUS or fluoroscopy may be completed to ensure adequate stent expansion and apposition against the arterial wall.

Designed to meet a spectrum of clinical challenges, Sprinter Legend RX offers:

  • 1.50 – 4.00 mm low profile balloons: The best options for treating a comprehensive spectrum of lesions
  • Micro-brite marker bands: Enhanced cross and track with uncompromised radiopacity
  • Broad product offering: 57 sizes, 10 diameters, including ¼ sizes and 7 lengths

And Sprinter Legend RX features Medtronic's renowned balloon technologies.

FasTrac tip

FasTrac Tip

With 0.41-mm lesion entry profile for superior crossability

MiniWrap folding

MiniWrap Folding

Proprietary process gives 1.50-4.00 mm balloons low-profile shoulder and tight wrap/rewrap

Dura-Trac selective coating

Dura-Trac Selective Coating

For excellent lubricity during tracking and position retention during inflation

Sprinter Legend 1.5 mm

Sprinter Legend RX Percutaneous Coronary Intervention

Sprinter Legend 3.0 mm

Sprinter Legend 3.0

 

Sprinter Legend RX Animation (3:23, 21.3 MB)
This video gives an overview of the Sprinter Legend RX catheter's high performance features.

Expand All

Specifications

Specifications

Specifications
Catheter length 142 cm
Distal shaft length 21 cm
Lesion entry profile (in.) 0.016
Tip length (mm) 2.50
1.50 mm crossing profile (in.) 0.021
Balloon material Fulcrum
Coating Selective DuraTrac
Platinum-iridium marker bands
  • 1.50 mm
  • All ODs >1.50 mm
Single
Double
Nominal pressure
  • 1.50 mm
  • All ODs >1.50 mm
6 atm
8 atm
Rated burst pressure
  • 1.50 mm
  • All ODs >1.50 mm
12 atm
14 atm
MiniWrap folds
  • 1.50 mm
  • 2.00 – 3.75 mm
  • 4.00 mm
2 folds
3 folds
5 folds
Shaft dimensions for 1.50 – 3.50 mm
  • Proximal
  • Distal
1.9 F
2.4 F / 2.6 F
Shaft dimensions for 3.75 – 4.00 mm
  • Proximal
  • Distal
1.9 F
2.5 F / 2.7 F
MGCID (in.) 0.056
Diameters (mm) 1.50 – 4.00
Lengths (mm) 6, 10, 12, 15, 20, 25, 30
Crossing Profile
Balloon diameter (mm) Crossing profile OD (in)
1.50 0.021
2.00 0.022
2.25 0.022
2.50 0.022
2.75 0.023
3.00 0.023
3.25 0.023
3.50 0.023
3.75 0.023
4.00 0.024

Compliance

Pressure
(atm)
1.5 2.00 2.25 2.50 2.75 3.00 3.25 3.50 3.75 4.00
4 1.36
5 1.41
6 1.47 1.91 2.23 2.34 2.60 2.81 3.14 3.32 3.61 3.85
7 1.53 1.95 2.29 2.41 2.67 2.89 3.22 3.42 3.70 3.96
8 1.58 2.00 2.33 2.47 2.74 2.97 3.29 3.50 3.79 4.05
9 1.61 2.03 2.38 2.53 2.79 3.04 3.35 3.57 3.86 4.12
10 1.64 2.06 2.41 2.57 2.84 3.09 3.40 3.63 3.93 4.18
11 1.67 2.09 2.44 2.61 2.89 3.14 3.44 3.68 3.98 4.23
12 1.70 2.12 2.47 2.64 2.93 3.18 3.48 3.73 4.02 4.29
13 1.73 2.14 2.51 2.67 2.96 3.22 3.52 3.77 4.07 4.35
14 1.75 2.16 2.54 2.70 3.00 3.25 3.56 3.82 4.12 4.40
15 1.78 2.19 2.58 2.73 3.03 3.29 3.59 3.87 4.16 4.45
16 2.21 2.61 2.76 3.07 3.33 3.64 3.92 4.21 4.50
Nominal pressure
Rated burst pressure*

Test data on file at Medtronic, Inc.
*Rated burst pressure: the maximum pressure to which a balloon is designed to be inflated (95% confidence that 99.9% will not fail below rated burst pressure). DO NOT EXCEED.
Nominal pressure: the pressure at which the balloon reaches its labeled diameter.

Ordering Information

References

1

Source from Medical Review: Caroline S. Rhoads, MD–Internal Medicine and Stephen Fort, MD, MCRP, FRCPC–Interventional Cardiology ©1995-2005, Healthwise.

IMPORTANT INFORMATION ON INDICATIONS, SAFETY AND WARNINGS

For listing of Indications, Contraindications, Warnings, Precautions, and Potential Adverse Events, please refer to the Instructions For Use .