placeEurope

Overview

Next Generation Single Channel LigaSure™ Generator

The new Valleylab™ LS10 electrosurgical generator provides you with the essential energy that you need.  Both affordable and portable, the LS10 generator makes the next generation of LigaSure™ vessel sealing technology accessible to customers around the globe. The platform is intended for open and laparoscopic surgical procedures and includes:

  • TissueFect™ sensing technology across all modalities
  • LigaSure™ fusion technology for vessel sealing

Using the Valleylab™ Exchange Software Update System, the unit can be easily updated on site, providing surgeons, nurses and patients with up-to-date, state-of-the-art technology. 

Specialties utilizing the Valleylab™ LS10 generator include: 

  • Colorectal 
  • Urology
  • General Surgery
  • Plastics
  • Gynecology

 

Ready to Set Up Your Valleylab™ LS10 Generator?

Find Out How
Order Information
Order Code Description Unit of Measure Quantity
VLLS10 Valleylab™ LS10 Generator Each 1
Specifications
General Characteristics
Output configuration Isolated output
Cooling Natural and forced convection
Display

System status indicator - circle LED (12 blocks) shows system status

  System error indicator - triangle LED with exclamation mark
  Instrument status indicator - LED above the LigaSure™ receptacle
  Usage limit indicator - 2 with a line through it
Mounting A Covidien cart (UC8009) or a stable, flat surface
Dimensions and Weight
Width 300 mm (11.81 inches)
Depth 377 mm (14.84 inches)
Height 105 mm (4.13 inches)
Weight 5 Kg (11 lbs)
LigaSure Output Characteristics
  Mode P-P Voltage Rated Load (Ohms) Maximum Power (Watts) Duty Cycle
LigaSure™ LigaSure™ 500 V 30 270 N/A
Input Power
100-120 Volt 220-240 Volt
  • Maximum VA – 400 VA
  • Input mains voltage, full regulation range 90-130 VAC
  • Input mains voltage, operating range 100-120 VAC
  • Mains line frequency range (nominal) 50 Hz to 60 Hz
  • Fuses (2) – 5 mm x 20 mm 8 A, 250 V fast blow
  • Maximum VA – 400 VA
  • Input mains voltage, full regulation range 180-264 VAC
  • Input mains voltage, operating range 210-240 VAC
  • Mains line frequency range (nominal) 50 Hz to 60 Hz
  • Fuses (2) – 5 mm x 20 mm 8 A, 250 V fast blow
 

Order Information